IEC 62368-1 Annex C UV Radiation Test Equipment , Xenon-Arc Light-Exposure Test
Product Details:
Place of Origin: | china |
Brand Name: | KingPo |
Certification: | ISO17025 |
Model Number: | KP-SW-280 |
Payment & Shipping Terms:
Minimum Order Quantity: | 1 |
---|---|
Price: | To be quoted |
Packaging Details: | safety carton pack or plywood box |
Delivery Time: | 35 working days |
Payment Terms: | TT |
Supply Ability: | 3 sets per month |
- Description
- Reviews (0)
Description
Detail Information
Place Of Origin::Guangdong, China (Mainland)Temperature Range:RT+10℃~70℃Humidity Range:Light 20% ~ 60%RH, Dark: 30% ~ 98%RHRain Time:0~9999min(adjustable)Rain Cycle:0~240min,adjustable IntervalIrradiated Intensity:(340nm) 0.35W/m²Lamp Life:1200hLamp Power:6500WHigh Light:
IEC 62368-1 UV radiation test equipment
,
Xenon Arc Light Exposure Tester
Product Description
IEC 62368-1 Annex C UV radiation test equipment , Xenon-arc light-exposure test
IEC62368-1 Standard requirement:
C.2 UV light conditioning test
C.2.1 Test apparatus
Samples are exposed to UV light by using one of the following apparatus:
– a twin enclosed carbon-arc (see C.2.3) with continuous exposure for a minimum of 720 h.
The test apparatus shall operate with a black-panel temperature of 63 °C ± 3 °C in a relative humidity of (50 ± 5) %; or
– a xenon-arc (see C.2.4) with continuous exposure for a minimum of 1 000 h. The test apparatus shall operate with a 6 500 W, water-cooled xenon-arc lamp, a spectral irradiance of 0,35 W/m² at 340 nm, a black-panel temperature of 63 °C ± 3 °C in a relative humidity of (50 ± 5) %
Specifications:
Item | Specification |
Internal Dimension (W*D*H) | 600*600*800mm(W*D*H) |
External Dimension (W*D*H) | 1350*1200*2050 mm(W*D*H) |
Inner chamber material | SUS 304 # Stainless Steel, Anti-rust, easy to clean |
Outer chamber material | cold plate rust proof and plastic spraying treatment |
Temperature | RT+10℃~70℃ |
Temperature Uniformity | ±3℃ |
Humidity Range | Light 20% ~ 60%RH, dark: 30% ~ 98%RH |
Irradiated intensity | (340nm) 0.35W/m² |
lamp power | 6500W |
Lamp life | 1200h |
Blank panel temperature | Light 60℃ ~ 100℃, dark 40 ~ 70℃ |
Temperature sensor | PT100 high precision temperature sensor |
Control system | PLC controller |
Distance from specimen to arc center | 250~300mm, Effective exposure area 3760cm² |
Filter | Borosilicate filter inside and outside |
Water pressure at nozzle | 0.12~0.15Mpa |
Nozzle diameter | 0.8mm |
Water supply time | 0-999min |
Cooling Model | Cold water machine |
Power | AC380V, 60Hz. |
Safety Protection | Over heat protection for fan, |
Over load protection for cooling system, | |
Over pressure protection for cooling system, | |
Over temperature protection, | |
Over heat and over flow protection for water pump, | |
Leakage protection, Water shortage protection, | |
Auto stop after failure warning |
Reviews
There are no reviews yet.