VOC Emission Test Chamber , Climatic Test Chambers 10±5pa Pressure
Product Details:
Place of Origin: | China |
Brand Name: | KingPo |
Certification: | Calibration Certificate |
Payment & Shipping Terms:
Minimum Order Quantity: | 1 |
---|---|
Price: | To be quoted |
Packaging Details: | safety carton pack or plywood box |
Delivery Time: | 35 working days |
Payment Terms: | T/T |
Supply Ability: | 3 sets per month |
- Description
- Reviews (0)
Description
Detail Information
ozone test chamber
,
climatic test chamber
Product Description
VOC Release Chamber
Usage:
* Confirm the influence of environment variables on the release rate and release properties of materials and products;
* Grade materials and products according to their release properties;
* Develop and validate the proposed model of indoor air pollutions;
* Evaluate indoor air quality(IAQ) according to the data and proposed model of indoor pollutants;
Implementation Standards:
* GB 18580 ''Indoor Decorating and Renovating Materials Limit of Formaldehyde Emission of Manmade Board and its Products''
* GB 18584 ''Indoor Decorating and Renovating Materials Limits of Harmful Substances from Wooden Furniture(Exposure draft)
* GB 18587 ''Indoor Decorating and Renovating Materials Releasing Limitation of Harmful Substances from Carpet,Carpet Pad and Carpet Adhesive''
* ENV717-1 "Environmental chamber method for measurement of formaldehyde emission of manmade board.
* EN13419-1 "Determination of VOCs Release quantity of building products Chapter One the method for releasing and testing the environmental chamber"
* HJ571-2010 "Technical requirements for environmental labeling manmade board and its products"
* HJ566-2010 "Technical requirements for environmental labeling manmade board and its products"
* ISO12219-4
* VDA-276
Reference Standards
* ASTM D5116-97 "Standard guide for determination of organic release from interior materials and products with small scale environmental chamber"
* ASTM D6330-98 "Standard operation for determination of VOCs(excluding formaldehyde) release from wooden board with small environmental chamber under specified test conditions"
* ASTM D6330-98 "Standard operation for determination of VOCs release from interior materials and products with full scale environmental chamber"
* ASTM D1333-96 "Standard test method for determination of formaldehyde concentration and release rate in gases released from wood products with large scale environmental chamber"
Technical Parameters:
Product Model |
V-50 |
V-100 |
V-250 |
V-500 |
V-1000A |
V-1000B |
V-2000 |
||||||
Volume of chamber(WxDxH)cm |
56*30*30 |
63*40*40 |
70*60*60 |
102*70*70 |
158*80*80 |
91*100*110 |
200*110*90 |
||||||
Environmental temperature |
5~35℃ |
||||||||||||
Environmental humidity |
≤90%R.H |
||||||||||||
Requirements for Environmental |
Cleanliness No pollutant source in the surrounding environment TVOC 500μg/m³ |
||||||||||||
Adjustable range of temperature |
10~40℃ high temperature cleaning 250℃ |
||||||||||||
Adjustable range of humidity |
40~70%R.H(20℃~30℃) |
||||||||||||
Pressure |
10±5pa |
||||||||||||
Temperature fluctuation |
±0.3℃(fluctuation) |
||||||||||||
Temperature deviation |
±0.5℃(deviation) |
||||||||||||
Humidity fluctuation |
±2.5%R.H |
||||||||||||
Humidity deviation |
±3% |
||||||||||||
Air exchange rate |
0.2~5/h±5%(optional 0.2~3/h±5%) |
||||||||||||
Air velocity |
0.1m/s 0.3m/s(Range) |
||||||||||||
Power |
Alternating current 380V/50Hz three-phase power supply 4/5 wiring system |
||||||||||||
Starting power |
5kw |
6kw |
8kw |
10kw |
15kw |
15kw |
20kw |
||||||
Operating power |
3kw |
3.4kw |
5kw |
6kw |
7kw |
7kw |
9kw |
||||||
Rate of recovery |
Formaldehyde,octanol,4 phenyl cyclohexene,dodecane and tetradecane have a recovery rate larger than 80% |
||||||||||||
Clean air system |
Standard configuration |
||||||||||||
Background concentration of methanol |
single VOC has a concentration smaller than 2 μg/m³,T VOC 20 μg m dust filtrated TVOC has a concentration smaller than 20μg/m³
|
Reviews